I N S T R U C T I O N
M A N U A L
TROUBLE SHOOTING -CONTINUED
Problem
Maintenance Mode message
Control unit blank
R E V
4 . 2
Possible Cause
Maintenance switch or
maintenance function is on
1- Control unit zero/span key
was pressed.
2- Service module
zero/operate switch in zero,
span/operate switch in span
3- Service module
test/operate switch in test
Control unit fuse open
47
Remedy
Return
all
to
operate
positions.
Replace and check for shorts
in the power supply or
individual boards.