2)Use the " "
Button " ( 3-6, Fig. 1 ) to adjust the desiring " High
Limit " value.
* When adjust the value, the " SV indicator " ( 3-3, Fig. 1 )
will light.
After adjust the " High Limit " value, press the " Set
Button " ( 3-4, Fig. 1 ) twice, " Display " will return to
the normal measuring screen.
4-6 2nd layer setting procedures
tPty
tPSt
HySt
tPoA
PCPS
Temperature unit setting
1)Press the " Set Button " ( 3-4, Fig. 1 ) continuously at
least two seconds, the " Display " will show " tPty ",
now the meter is ready for the Temperature unit (
) setting.
℉
2)Use the " "
Button " ( 3-6, Fig. 1 ) to adjust the desiring
temperature unit to " C " or " F ".
* When adjust the Temp. unit, the " SV indicator " ( 3-3, Fig. 1 )
will light.
Button " ( 3-5, Fig. 1 ) and the "
▼
Temp. unit setting
Temp. compensation
value setting
Hysteresis setting
Temp. Offset setting
Probe Calibration
Period Setting
Button " ( 3-5, Fig. 1 ) and the "
▼