System Check: Control Unit And Logger; Troubleshooting - TESTO 454 Getting Started

And control unit
Hide thumbs Also See for 454:
Table of Contents

Advertisement

Available languages

Available languages

4.6 System check: Control unit and logger

4.6.1 Connecting probes
Probes are only recognised by the Control unit and the logger if they were already
connected prior to switch-on.
Insert the probe into one of the four sockets available in the logger.
4.6.2 Switching on Control unit
Press
to switch on Control unit.
- After a brief initialisation phase, the Control unit measurement menu switches
automatically to the measurement view in the logger.
The readings from the connected probes are shown in the order of the probe socket
allocation.
With more than six readings:
Use
to scroll between the displays.
- The current display page is shown in the top display line on the right.
Attach a terminal plug to the last instrument (logger).
The following must be available after the Control unit is switched on:
1 Status LED on logger lights up green.
2 Control unit display lights up.
3 The readings from the probes connected to the logger are shown in the display.

Troubleshooting

Fault
LED on logger does not light up.
Control unit cannot be switched on.
Probes are not recognised.
Loggers are not recognised
4.6.3 Switching off Control Unit and logger
Press
to switch off Control unit.
Cause
No power in logger.
Rechargeable battery pack is empty.
No power to Control unit.
Rechargeable battery pack is empty.
Probes are not connected correctly.
Probes are defective.
No databus connection
4. Initial operation 43
Remedy
Check if the instruments are
connected correctly to each
other:
· Databus cable/ power cable
· Contact strips
Recharge rechargeable battery.
Check if the instruments are
connected correctly to each
other:
· Databus cable/power cable
· Contact strips
Recharge rechargeable battery.
Connect probes correctly.
Change probes.
Check bus connection.
(terminal plug)