USB 2.0 Limitations
The following information covers the currently known limitations associated with
operating under the USB 2.0 interface.
Demo Port Capability
DIF
Kinetics
Reset Camera to NVRAM
Defaults
Temperature Lock Status
PTG
Virtual Chip
Custom Timing
Custom Chip
Frames per Interrupt
RS170 (Video Output)
Online Exposure
File Information
Overlapping ROIs
Maximum cable length is 5 meters (16.4 feet)
1 MHz is currently the upper digitization rate limit for the ST-133 Controller.
Large data sets and/or long acquisition times may be subject to data overrun
because of host computer interrupts during data acquisition.
USB 2.0 is not supported by the Princeton Instruments PC Interface Library (EZ-
DLLS).
Some WinX (WinView/32 and WinSpec/32) 2.5.X features are not fully
supported with USB 2.0. See the table below.
Feature
Table 15. Features Supported under USB 2.0 (continued on next page)
Supported with USB 2.0 in
WinX 2.5.X
NO
YES
NO
YES
YES
NO
YES
YES
NO
NO
NO
YES
NO
Appendix G
Remarks
WinX 2.5.18.1
WinX 2.5.x doesn't utilize
hardware lock status
WinX 2.5.18.1
WinX 2.5.18.1
Not all header info is
currently available in
WinX 2.5.x through
PVCAM
217
Need help?
Do you have a question about the PI-MAX System and is the answer not in the manual?
Questions and answers