5 Setup
5.2

Setup steps

Parameter data
After transmission of the sensor data, all application-specific settings
must be entered again. Carry out a fresh setup after exchanging the
electronics or load the stored data of the setup.
If you saved the parameter settings during the first setup of the sen-
sor, you can transfer them to the replacement electronics module. To
do this, use the import function of the adjustment software PACTware
with the device DTMs or the copy function of the display and adjust-
ment module. A fresh setup is then not necessary.
You can find information on how to copy saved parameter adjustment
data to the new electronics module under "Additional settings - Copy
instrument settings".
On SIL qualified instruments, the settings of the electronics must be
checked and verified after the electronics exchange. Only then is the
instrument again ready for operation.
You can find information about the test and verification in the operat-
ing instructions corresponding to your sensor as well as in the Safety
Manual.
Information:
If a false signal suppression already existed, we recommend creating
a new one. If this is not possible, for example, because the vessel
is filled, the false signal suppression can be imported via the DTM.
However, the false signal suppression should be updated or created
anew next time the vessel is empty.
During setup, take note of the operating instructions corresponding to
your sensor.