Setup Steps - Vega VEGACAL 65 Operating Instructions Manual

Capacitive cable electrode for continuous level measurement
Hide thumbs Also See for VEGACAL 65:
Table of Contents

Advertisement

Switch-on phase
Parameter adjustment
Carry out min. adjustment
VEGACAL 65 • Foundation Fieldbus
6 Set up with the display and adjustment module PLICSCOM
Approx. 60 minutes after the last pressing of a key, an automatic reset
to measured value indication is triggered. Any values not confirmed
with [OK] will not be saved.
6.4

Setup steps

After VEGACAL 65 is connected to voltage supply or after voltage
recurrence, the instrument carries out a self-check for approx. 30
seconds. The following steps are carried out:
Internal check of the electronics
Indication of the instrument type, the firmware as well as the sen-
sor TAGs (sensor designation)
Status byte goes briefly to fault value
Then the current measured value will be displayed and the corre-
sponding digital output signal will be output to the cable.
VEGACAL 65 measures the capacitance of the respective product. To
display the actual level of the product, an allocation of the measured
capacitance to the percentage height must be carried out. For this
adjustment, the capacitance is entered with emptied and filled vessel.
If the vessel cannot be emptied or filled completely, you can carry out
the adjustment also with two known levels - for example with 10 %
and 90 %. The difference between the empty and full adjustment
values should be as large as possible.
The actual level can then be calculated on the basis of these settings.
VEGACAL 65 must be installed. A change of level is necessary for
this adjustment.
In the main menu item " Basic adjustment", the individual submenu
items should be selected one after the other and provided with the
correct parameter values.
Tip:
If the display and adjustment module remains on the probe as a
display, we recommend saving the sensor data in the display and
adjustment module.
Use the function "Copy sensor data".
Start your parameter adjustment with the following menu items of the
basic adjustment:
To be on the safe side, note the adjustment values for full and empty.
If an adjustment procedure fails, it is not necessary to again carry out
a level change.
These values can be helpful if the electronics has to be exchanged.
Empty adjustment
Full adjustment
The values correspond to the actual measured level as well as to the settings
1)
already carried out, e.g. default setting.
%
1)
Value
27

Advertisement

Table of Contents
loading

Table of Contents