8
Start/Stop ventilation button
9
Main screen
10
Sub-menus
11
Pressure bar
Note: Do not access Clinical mode

Information bar

The Information bar is displayed at the top of the touch screen. The Information bar displays the
operating status of the device, including patient type, current circuit configuration, programs, information
messages, ventilation status, alarms and power status.
Description
Patient type – Adult
Patient type – Paediatric
Circuit type – Single limb with intentional leak
Circuit type – Single limb with expiratory valve
Circuit type – Double limb
P1
Program number and ventilation mode in use
(A)CV
Multiple alarms are active simultaneously. The highest priority active alarm is displayed first.
Message
Will display alarms or information. Image above shows device in Standby. (Displayed when the device is powered
window
on but not ventilating). Date and time will be displayed when the device is ventilating and there are no active
alarms.
Information messages are displayed in blue text. If the device Alert tone setting is 'On', you will be alerted to new
information messages by a single beep.
unless directed by a clinician.