Table 3:
Approved cleaners
Product
PI-Spray II
a
For maximum effectiveness, the component being cleaned must remain wet with disinfectant for a minimum period
of time.
b
Qualified for use as an intermediate-level disinfectant for mycobacteria.
Refer to the cleaners and disinfection document available at
current cleaners and disinfectants.
Table 4:
High-level compatible disinfectants
Disinfectant
Compatible transducers
Cidex
C8x, C11x, C35x, HFL38xi,
HFL50x, HSL25x, ICTx, L25x,
L38xi, P10x, P11x, rC60xi, rP19x
Cidex OPA
C8x, C11x, HFL38xi, HFL50x,
HSL25x, ICTx, L25x, L38xi, L52x,
P10x, rC60xi, rP19x
Revital-OX
C35x
Resert
Refer to the cleaners and disinfection document available at
current cleaners and disinfectants.
Cleaning and disinfecting
Compatible Transducer
C8x, C11x, HFL38xi, HFL50x,
HSL25x, ICTx, L25x, L52x, L38xi,
P10x, P11x, rC60xi, rP19x
Minimum wet contact time
10 minutes
www.sonosite.com
for a complete list of the most
Temperature
Disinfectant Soak Duration
o
o
45 minutes
25
C, 77
F
o
o
12 minutes
20
C, 68
F
o
o
8 minutes
20
C, 68
F