Weight On Demand Formats - Cardinal MED VUE Mv1 Owner's Manual

Medical weight analyzer
Table of Contents

Advertisement

WEIGHT ON DEMAND FORMATS

Weight On Demand
If the continuous output was not selected for the serial port, RS232 MOdE=0, the USB Device
port, USb MOdE=0 or the optional Ethernet/WiFi port, ETHERNET MOdE=0 and the indicator is
connected to a computer for transmission of weight data to a PC-based EMR (electronic
medical record) software program, it will transmit a single set of weight data each time the
computer sends an weight request "ENQ" (hex 05) or a SMA weight request "W". This is
known as Weight-On-Demand. Examples and explanation of the data format transmitted are
shown below.
<ENQ> - (Hex 05)
Character sent to the MedVue weight analyzer will respond with the following weight string:
Pxxxxxx^UU^M^SS^CR (no decimal point in weight)
PxxxxxxD^UU^M^SS^CR (decimal point in weight)
Where:
P = polarity (space if positive, - if negative)
xxxxxx = weight with leading spaces
D = decimal point (embedded where necessary)
^ = space (hex 20)
UU = units, upper case (LB, KG, TN, etc.)
M = mode, upper case (G=gross, N=Net)
SS = status, upper case (CZ=center-of-zero,
MO=motion, BZ=gross weight below zero
EE=entry - Input in progress, OC=over capacity)
CR = carriage return (hex 0D)
<LF>W<CR>
String sent to the MedVue weight analyzer will respond with the SMA weight string:
<LF>S1GM^DDDDDDDDDDUUU<CR>
Where:
<LF> = Line feed character (hex 0A)
S = Status ('O' = Over Cap, 'Z' = Center Zero, 'U' = Below Zero, 'E' = Error)
1 = The number '1'
G = Mode of operation ('G' = Gross, 'N' = Net, 'T' = Tare)
M = Motion bit ('M' = Motion, ' ' = Settled)
^ = Space
DDDDDDDDDD = Weight with decimal point if necessary
UUU = Units (e.g. 'lb ', 'kg ', 'ton', etc..)
<CR> = Carriage return (hex 0D)