bARCOdE (Use ID Scanner)
This setting determines whether the MV1 allows a scanner to be used to input an ID number
to be entered prior to beginning the weighing operation. This applies to both Basic
Operation and with the BMI feature enabled.
The display will show bARCOdE and the current setting. If the value displayed is acceptable,
press the PRINT/ENTER key to save it. Otherwise use the numeric keys 0/NO or 1/YES
and enter a new setting. Press the PRINT/ENTER key to save it.
Allowable values are YES or no.
YES = ID Scanner is used to input ID
number
USE
H EIGHT (Allow Height Functions)
Z
This setting determines whether the height functions are allowed and if the Height Units
prompt is shown.
The display will show USE HEIGHT and the current setting. If the value displayed is
acceptable, press the PRINT/ENTER key to save it. Otherwise use the numeric keys 0/NO
or 1/YES and enter a new setting. Press the PRINT/ENTER key to save it.
Allowable values are YES or no.
YES = The Height functions are
allowed and the HEIGHT UNITS
prompt will be displayed
NOTE: If USE HEIGHT = no (no) is selected, the time will be shown on the display in the
location where the height normally would be.
HEIGHT UNITS (Height Units –
The display will show HEIGHT UNITS and the current setting. If the value displayed is
acceptable, press the PRINT/ENTER key to save it. Otherwise, using the numeric keys,
enter a new value and then press the PRINT/ENTER key to save it.
Allowable values are 1 or 2.
1 = Centimeters
bOdY MASS IdX (Body Mass Index, BMI Calculation Feature)
This setting determines whether the MV1 performs basic weigh calculation only or performs
the Body Mass Index (BMI) calculations.
The display will show bOdY MASS IdX and the current setting. If the value displayed is
acceptable, press the PRINT/ENTER key to save it. Otherwise use the numeric keys 0/NO
or 1/YES and enter a new setting. Press the PRINT/ENTER key to save it.
Allowable values are YES or no.
YES = Body Mass Index is Enabled
and MV1 performs BMI calculations
8555-M512-O1 Rev D 01/15
no = ID Scanner is not used
no = The Height functions are not
allowed. Proceed to the
HI RESOLUTION parameter
Only if USE HEIGHT = YES and UNITS > 1
2 = Millimeters
no = Body Mass Index is Disabled and
the MV1 performs basic weight
calculations only
)