Tare Weight Entry (Tare weight is known)
1. In Gross Weight mode (GROSS annunciator on), press TARE key.
2. Display will show )0 or previously entered tare and TARE annunciator will turn on.
3. If weight displayed is acceptable, press PRINT/ENTER key to save it. Otherwise, use
the numeric keys and enter a new tare weight. Note that when entering tare weights, a
maximum of 4 digits can be entered and the numbers advance from right to left in the
display. The number of leading or trailing zeros required to obtain your desired tare is
dependent upon the DIVISION VALUE selected in the setup procedure.
For example:
4. After the new tare weight has been entered, press PRINT/ENTER key.
5. Display will show a minus Net weight and NET annunciator will turn on.
6. Press ID/HEIGHT key.
7. Display will change to show Id.
8. Using numeric keys, enter up to an 14 digit alpha numeric identification number.
9. Press PRINT/ENTER key.
10. Place patient on scale
11. Read weight display.
12. If a printer is connected to scale, press PRINT/ENTER key to print a ticket.
13. Remove patient from scale.
8555-M512-O1 Rev D 01/15
Division Value:
Desired Tare Value:
Key Sequence:
0.2 lb
10.6 lb
1 0 6