CALIbRATION
With CALIbRATION displayed, press the PRINT/ENTER key. The display will change to
CALIbRATE. Proceed to the Perform Calibration parameter.
Otherwise, press the ID/HEIGHT key to advance to the GRAVITY menu.
CALIbRATE (Perform Calibration)
The display will show CALIbRATE and the current setting (NO).
If the scale has been previously calibrated and you wish to skip calibration (and retain
the internal calibration factor), press the PRINT/ENTER key. The display will change
to GRAVITY ENAbLE. Proceed to the GRAVITY menu, Gravity Enable parameter.
If calibration is desired, press the 1/YES key to select YES and then press the
PRINT/ENTER key. The display will change to NUM CAL POINTS.
Allowable values are YES or no.
YES = Perform Calibration and
proceed to NUM CAL POINTS
NUM CAL POINTS (Number Of Calibration Points –
The MedVue Model MV1 Medical Weight Analyzer has two (2) modes that can be used to
perform calibration. The calibration modes are as follows:
Single Calibration Point
This is a standard calibration method requiring one weight, an empty scale and has one
conversion factor. This method uses two calibration points (a test load on the scale and
an empty scale) to establish a zero (no load) calibration value and to span the MV1.
Multi-point Calibration
This method requires two weights, an empty scale and has two conversion factors. It
uses three calibration points, (a load on the scale, half the previous load and an empty
scale). The three points correspond to zero weight, mid-point weight and test load
(weight).
The display will show NUM CAL POINTS and the current setting. If the value displayed is
acceptable, press the PRINT/ENTER key to save it. Otherwise, using the numeric keys,
enter a new value and then press the PRINT/ENTER key to save it.
Allowable values are 1 or 2.
1 = Single Calibration Point
8555-M512-O1 Rev D 01/15
no = Skip Calibration and proceed to
the GRAVITY ENAbLE parameter
Only if CALIbRATE = YES
2 = Multi-Point Calibration
)
27