Essential Measurement Practices

Agilent 2100 Bioanalyzer System

Agilent 2100 Bioanalyzer System
• Do not touch the 2100 Bioanalyzer instrument during a run and never place
• Do not force the chip to fit in the 2100 Bioanalyzer instrument. The
• Clean electrodes on a daily basis using the electrode cleaner. For more
• Thorougly clean electrodes on a monthly basis using a toothbrush and
• Clean the focusing lens once a month (or after any liquid spill) using
14
it on a vibrating surface or near air-circulating instruments (for example,
temperature cyclers).
electrode cartridge may be damaged when the lid is closed.
details, see
"Maintenance of the Electrode Cartridge"
distilled water. For more details, see
Electrode Cartridge"
isopropanol. For more details, see
"How to Clean the Pin Set of the
on page 126.
"Cleaning the Lens"
on page 115.
on page 143.
Maintenance and Troubleshooting