Status Listings

OSA Notices

OSA Notices

System control-related error messages or warnings
The OSA system changed a setting and generated a warning that the opera-
tion was performed.
Table 5-5. System Control Errors or Warnings
Error Number
Error Description [description/explanation/examples]
1000
desc = "Sensitivity forced to Auto"
help = "Sensitivity has been forced to the Auto setting because another
instrument setting was made which does not permit sensitivity to be in the
Manual setting."
1001
desc = "Old Mezzanine board in system"
help = "This OSA has an old mezzanine board installed. It is not necessary to
upgrade the mezzanine board but the software will not be able to correctly
determine which light source options are available."
1002
desc = "No Delta Marker Amplitude for"
help = "The delta marker amplitude display has been turned off due to a units
mismatch. The amplitude units of the reference marker are different than those
of the delta marker. Values in differing units cannot be compared."
1003
desc = "Trace length increased"
help = "Trace length has been increased because the wavelength limit function
was turned off. The minimum trace length is larger when the wavelength limit
function is off and trace length was below the new minimum value."
2999
desc = "The Notices list has overflowed"
help = "The Notices list has overflowed. The last entries received have been
deleted."