HP 86140A Series User Manual page 191

Optical spectrum analyzer
Table of Contents

Advertisement

and
[SENSe:][WAVelength:]CENTer:STEP:AUTO OFF|ON|0|1
[SENSe:][WAVelength:]CENTer:STEP:AUTO?
When
value is set by the [SENSe:][WAVelength:]CENTer:STEP[:INCRement] com-
mand.
[SENSe:][WAVelength:]CENTer:STEP[:INCRement] <numeric_value>[M|NM|UM|A]
[SENSe:][WAVelength:]CENTer:STEP[:INCRement]?
Specifies the center wavelength step size.
[SENSe:][WAVelength:]LIMit OFF|ON|0|1
[SENSe:][WAVelength:]LIMit?
Specifies whether the span is limited to the specified range of 600 to 1700 nm.
[SENSe:][WAVelength:]OFFSet <numeric_value>[M|NM|UM|A]
[SENSe:][WAVelength:]OFFSet?
Specifies the wavelength offset. This is the offset between the measured
wavelength and the displayed wavelength.
Span
Center
=
Stop
------------- - """"
2
With Wavelength Limit Off, the minimum value for the Center Wavelength is nominally
350.1 nm. The maximum value is 1999.9 nm. These limits are valid for wavelengths ref-
erenced in air or vacuum.
With Wavelength Limit On, the minimum value for the Center Wavelength is nominally
600.1 nm. The maximum value is 1699.9 nm. These limits are valid for wavelengths ref-
erenced in air or vacuum.
The Preset value for Wavelength Limit is On. The Preset value for Center Wavelength is
1150 nm.
, the step size is automatic. When
ON
SENSe Subsystem Commands
, the step size is fixed. The
OFF
Remote Operation
4-85