EasyGuide

1 Introduction

Thank you for choosing the EasyOne
With EasyOne
you have chosen a high quality spirometer that minimizes the need for maintenance
due to its unique ultrasound technology. EasyOne
ly accurate over years. The spirette
tient at low cost even if the spirometer is frequently used.
The EasyOne
Diagnostic Spirometer has two operating modes for you to choose from:
In the Diagnostic mode, EasyOne
accordance with the standards of the European Respiratory Society (ERS) and the American Thoracic
Society (ATS).
In the Frontline mode, EasyOne
In the NLHEP mode, the EasyOne
gram (NLHEP [4]). This mode is a little more restrictive than the Frontline Mode (only FEV6 maneuvers).
The EasyOne
Frontline Spirometer only offers the Frontline and NLHEP mode. The differences be-
tween the two operating modes are described in the table that follows.
Test modes
Parameters
Quality control
Automatic quality con-
trol
Trial storage and dis-
play
Report Configuration
The default setting of the EasyOne
the Frontline Mode, see Chapter 8 under General Settings .
The EasyOne -line Spirometer does not contain the EasyOne
EasyOne
Screen Connector in conjunction with the EasyWare software for the PC. The screen connec-
tor can be used to display real time curves on the PC Screen, it can however not be used for direct con-
nection of EasyOne
ware.
Spirometer.
breathing tube assures perfectly hygienic conditions for every pa-
offers you extensive and diverse options for spirometric tests in
offers you the option of greatly simplified spirometric measurement.
fulfills all requirements of the National Lung Health Education Pro-
Diagnostic mode
FVC (expiratory), F/V Loop (inspiratory
and expiratory), slow VC, MVV, Pre-
Post measurement
US: Unit can be configured to meet
NIOSH/OSHA and Disability reporting
requirements.
FEV1, FVC, FEV1/FVC, FEV6,
FEV1/FEV6, MEF25-75, MEF25, MEF50,
MEF75, PEF, FET, FIVC, PIF, IVC, IRV,
ERV, FEV1/VC, MVV, pre-post % varia-
tion, QC rating
Requires 3 acceptable, reproducible
maneuvers.
Details in Chapter 10.1.
Quality control can also be overridden
manually.
Can store and display the best, or the
best 3 trials, including curves.
Report can be customized for curve
type and size.
Spirometer is the Diagnostic Mode. To switch the EasyOne
to a printer. The EasyWare manual describes installation and use of the PC soft-
does not need calibration and remains consistent-
Frontline and NLHEP mode
FVC (expiratory), Pre-Post measurement
FEV1, FEV6, FEV1/FEV6, FVC, FEV1/FVC,
PEF, pre-post % variation, QC rating
NLHEP mode: only FEV6, no PEF display
Requires 2 acceptable, reproducible ma-
neuvers.
Details in Chapter 10
Automatic control is always active
Stores and displays only the best trial and
curve
Report is fixed, showing the smaller sized
FV and VT curves
cradle. Instead it contains the