Outputting The Data Storage - wtw pH 1970i Operating Manual

Portable ph measuring instrument
Table of Contents

Advertisement

pH 1970i
Outputting stored data
on the display
ba75690e01
01/2007

4.4.3 Outputting the data storage

You can output the contents of the data storage:
Stored data on the display
Calibration data on the display
Stored data on the serial interface
Calibration protocol on the interface
1
Press the <RCL> key repeatedly until StO dISP appears on the
display.
pH
S
I
S T
O
Sal
Time
Baud
D
I
S
Day.Month
No.
Year
Ident
Tref25
Tref20
Auto
Store
LoBat
nLF
Lin
Oxi
Cal
TEC
2
Press the <RUN/ENTER> key.
A measured value appears on the display.
The storage location of the data record is displayed for approx.
2 s, then the respective temperature appears.
pH
S
6
8 7
1
O
Sal
Time
Baud
8
2 4 8
Day.Month
No.
Year
Ident
Tref25
Tref20
Auto
Store
LoBat
nLF
Lin
Oxi
Cal
TEC
You can perform the following activities:
Display further elements of the data record
(ID number, date, time, storage location)
Advance one data record (storage location)
Go back one data record (storage location)
mV
/
pH
o
%
mg/l
S/
m
cm
1/
cm
Sal
P
C
°
/
K
%
TP
REL1
REL2
ARng
AR
RCL
mV
/
pH
% mg/l
S/
m
cm
1/
cm
Sal
C
°
/
%
K
TP
REL1
REL2
ARng
AR
RCL
Operation
Press
<RUN/ENTER>
Press <L>
Press <M>
37