Safety Information

In this manual, a Warning identifies conditions and actions that pose hazard(s)
to the user. A Caution identifies conditions and actions that may damage the
Logger or the test instruments.
800.561.8187
To prevent possible electrical shock or personal
injury, follow these guidelines:
Follow all safety instructions.
Adhere to local and national safety codes.
Individual protective equipment must be used to
prevent shock injury where hazardous live
conductors are exposed.
The Logger may only be used and handled by
qualified personnel.
Remove all test leads from the Logger before you
open the battery door. Open the Logger only to
replace the rechargeable battery.
Maintenance work must be done only by qualified
service personnel.
Use only specified current probes. If you use
flexible current probes, wear suitable protective
gloves or work on de-energized conductors.
Protect the Logger against dampness, wetness,
and humidity.
Always connect voltage and current test leads to
the Logger before connecting to the load.
The plug and socket connection for the voltage
lead set is designed for 600 V CAT III. The
maximum voltage between outer conductor and
earth potential must not exceed 600 V. With multi-
phase connections, phase-phase voltage may not
exceed 800 V.
Use only the provided original or specified
accessories. This includes the ac power adapter.