Adjusting the centering spring
We recommend that you make these adjustments using the 54-vial tray. This is especially
N O T E
important if you intend to use the 54-vial tray in your work.
1 Open the front door of the Well-plate Sampler.
2 Remove the plate from the transport position of the Automation Interface
3 Check that the plate slides easily left and right.
Agilent 1260 Infinity Sample Capacity Extension User Manual
and place it in the rear position in the Well-plate Sampler.
If the plate does not slide easily, remove it and compress slightly the
centering spring on the rear wall. Recheck the movement.
Installation
2
53