5.4.9 Calibration dialog

In the CALIBRATION DIALOG menu you can
enhance or restore touch screen response to
buttons being pressed. We recommend to use a
pen for the calibration procedure.
1
3
Leica ST5020
• PRINT button:
Press for a printout of
the current station con-
figuration.
Instead of the graphical screen dis-
play, a list
which, in addition to the information
provided by the graphical screen, also
shows the reagents allocated to each
station.
Fig. 47
The pen must be placed exactly on the red dot
(1). The dot will move as shown in
exit the menu, press the dot once more in Posi-
tion 4.
Functional test:
Calibrate the screen and verify the result ap-
proaching one of the screen buttons from all
four sides - if the screen has been successfully
calibrated, the button may not appear as pressed
(wider border) before you actually touch it.
Standard appearance
2
Fig. 48
5.
Operation
(Fig.
47) will be printed,
Fig.
29. To
Button pressed
51