5.4.9 Calibration dialog
In the CALIBRATION DIALOG menu you can
enhance or restore touch screen response to
buttons being pressed. We recommend to use a
pen for the calibration procedure.
1
3
Leica ST5020
• PRINT button:
Press for a printout of
the current station con-
figuration.
Instead of the graphical screen dis-
play, a list
which, in addition to the information
provided by the graphical screen, also
shows the reagents allocated to each
station.
Fig. 47
→
The pen must be placed exactly on the red dot
(1). The dot will move as shown in
exit the menu, press the dot once more in Posi-
tion 4.
Functional test:
Calibrate the screen and verify the result ap-
proaching one of the screen buttons from all
four sides - if the screen has been successfully
calibrated, the button may not appear as pressed
(wider border) before you actually touch it.
Standard appearance
2
Fig. 48
5.
Operation
(Fig.
47) will be printed,
→
Fig.
29. To
Button pressed
51