Table 12.4. Immunity Test Table - biodex GAIT TRAINER 3 Application/Operation Manual

Hide thumbs Also See for GAIT TRAINER 3:
Table of Contents

Advertisement

Table 12.4.
Immunity Test Table
Immunity
Test
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air Test level
IEC 61000-4-2
Electrical fast
± 2 kV for power
transients/
lines
burst
± 1 kV for
IEC 61000-4-4
input/output lines
Surge
± 1 kV differential
IEC 61000-4-5
mode
± 2 kV common
mode
Voltage dips,
< 5% UT (> 95% of
short
dip in UT for 1/2
interruptions
cycle)
and voltage
40% UT (60% of dip
variations on
in UT) for 5 cycle
power supply
70% UT (30% of dip
input lines
in UT) for 25 cycle
IEC 61000-4-11
< 5% UT (> 95% of
dip in UT) for 5 sec
Power
3 A/m
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Conducted RF
3 Vrms,
IEC 61000-4-6
150 KHz to 80 MHz
Conducted RF
3 Vrms,
IEC 61000-4-6
150 KHz to 80 MHz
GAIT TRAINER 3 (v3.X SOFTWARE)
IEC 60601-1-2
Test Level
Contact ± 6 kV
Air ± 8 kV
Power ± 2 kV
Signal ± 1 kV
± 1 kV diff. mode
± 2 kV com. mode
< 5% UT (> 95% of
dip in UT) for 1/2
cycle
40% UT (60% of dip
in UT) for 5 cycle
70% UT (30% of dip
in UT) for 25 cycle
< 5% UT (> 95% of
dip in UT) for 5 sec
3 A/m
3 Vrms,
150 KHz to 80 MHz
3 Vrms,
150 KHz to 80 MHz
IEC 60601-1-2
Compliance
Level
Floor should be wood,
concrete or ceramic tiles.
If floor is covered with
synthetic material, the relative
humidity should be at least
30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If a
better mains power quality is
required, it is recommended
that the Gait Trainer 3 is
powered from an
uninterruptible power supply.
If image distortion occurs, it
may be necessary to position
the Gait Trainer 3 display
further from sources of power
frequency magnetic fields or to
install magnetic shielding. The
power frequency magnetic
field should be measured in
the intended installation
location to assure that it is
sufficiently low.
Portable and mobile RF
communications equipment
should be used no closer to
any part of the Gait Trainer 3,
including cables, than the
recommended separation
Electromagnetic
Environment -Guidance
103