12. Electromagnetic Compatibility
Conformance to Standards
This equipment conforms to the following safety standards:
Table 12.1.
Standard
IEC60601-1-2
Accompanying EMC Documents
This medical electrical equipment needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information provided in this manual.
!
Portable and mobile RF communications equipment can affect medical electrical
equipment.
!
Use of accessories, transducers, and cables other than those specified, with the
exception of accessories, transducers, and cables sold by the manufacturer of this
equipment, as replacement parts for internal and external components, may result in
increased emissions or decreased immunity of the equipment.
!
The Gait Trainer 3 should not be used adjacent to or stacked with other equipment. If
the Gait Trainer 3 is used while positioned adjacent to other equipment, it should be
observed to verify normal operation in the configuration in which it will be used.
List of Cable Accessories
The table below includes all accessory cables supplied with the Gait Trainer 3 for which the
manufacturer of this equipment claims compliance to EN 60601-1-2 when used with the Gait
Trainer 3.
Table 12.2.
Cable Accessories Table
Cable Description
USB Printer Cable
GAIT TRAINER 3 (v3.X SOFTWARE)
Safety Standards Conformance Table
Edition and/or Date
First edition, 2007
Part Number
Biodex # C12086
Cable Length
15ft