Error
Description of Error
Code
E55
Buffer Overflow
E56
Motor Current High While Blower
On
E57
Unable to Queue Data to
Provide Therapy
E58
Stuck Key
E59
Outlet Pressure Sensor Railed
E60
Blower Pressure Sensor Railed
E61
Real-time Clock's Battery Dead
E62
Blower Speed exceeds
maximum limit
E63
Internal Watchdog Failure
E64
External Watchdog failure
E65
Unexpected Watchdog Reset
E66
Reserved for Future Use
E67
Operating System Failed
Initialization
E68
Unable to Queue Data for
Communication
E69
Sampling Thread Locked
E70
Execution Thread Locked
E71
Internal RAM Failure
E72
Unable to Queue Data for
logging
E73
Reserved for Future Use
1032443, R
. 00
EV
B
PAP S
I
YNCHRONY
S
& T
ERVICE
ECHNICAL
Probable Cause
Defective PCA
Stalled or Locked Blower
Defective Motor Drive Chip
Defective FLASH or RAM
chip
Verify all keys protrude
through top
Defective Keypad
Defective Pad on PCA
Occluded or restricted
pressure tube
Defective pressure sensor
Occluded or restricted
pressure tube
Defective pressure sensor
Dead Battery
Defective Motor Drive chip
Defective RAM chip
Defective Microprocessor
Defective FLASH or RAM
chip
Defective RAM chip
Defective FLASH or RAM
chip
Defective FLASH or RAM
chip
Defective PCA
Defective PCA
Defective Microprocessor
Defective FLASH or RAM
chip
Defective RAM chip
T
ROUBLESHOOTING
I
NFORMATION
Corrective Action
Replace Main PCA
Replace Blower
Replace Main PCA
Replace Main PCA
Replace Main PCA
Align keys with top
Replace Keypad
Replace Main PCA
Replace Pressure
Tube
Replace Main PCA
Replace Pressure
Tube
Replace Main PCA
Replace Main PCA
Replace Main PCA
Replace Main PCA
Replace Main PCA
Replace Main PCA
Replace Main PCA
Replace Main PCA
Replace Main PCA
Replace Main PCA
Replace Main PCA
Replace Main PCA
Replace Main PCA
& D