14. The program will then run through a series of humidifier (optional), lighting, alarm, and dis-
play tests. Follow the on screen instructions.
15. When the prompt in Figure N appears, the program is ready to start sensor calibration.
Click on the "OK" button.
16. Observe the manometer. If the pressure is higher than 4.00 cm H
ton to decrease the blower speed. If the pressure on the manometer is lower than 4.00
cm H
O, click on the green button to increase blower speed. Each time a button is
2
clicked, the pressure will change by approximately 0.2 cm H
achieved, click "DONE" on the screen.
Let the pressures stabilize for at least one minute before making any adjustments
during the calibration.
1032447, R
. 00
EV
B
PAP S
I
YNCHRONY
F
M: H
IGURE
UMIDIFIER
F
N: S
IGURE
YSTEM
NOTE
S
& T
ERVICE
ECHNICAL
T
S
EST
CREEN
C
ALIBRATION
O. When 4.00 cm H
2
T
ESTING
I
NFORMATION
O, click on the red but-
2
O is
2
& C
- P
ALIBRATION
AGE
11