Respironics BiPAP Synchrony Service And Technical Reference Manual page 62

Hide thumbs Also See for BiPAP Synchrony:
Table of Contents

Advertisement

Error Code
E25
Excessive Drift on Flow Sensor
E26
Communications Failure with LCD Driver Chip
E27
Excessive Drift on Outlet Pressure Sensor
E28
Empty Calibration Table
E29
Excessive Drift on Blower Pressure Sensor
E30
Unrecognized Version of Parameter Storage
E31
Un-repairable Parameter Storage
E32
Corrupt Parameter Storage
E33
Excessive Parameter Storage Size
E34
Unable to Queue Data to Parameter Storage
E35
Unable to write to parameter storage
E36
Parameter Out of Range
E37
Corrupt Real-Time Clock Value
E38
Real-Time Clock not Ticking
E39
Unable to Queue Data to User Interface
E40
Invalid Built-in Self Test Call
E41
Over Pressure Condition
E42
Operating System not Responding to Software
E43
Insufficient Voltage for Audible Alarm
E44
12V Reference Out of Range
E45
5V Reference Out of Range
E46
Bulk Voltage Out of Range
E47
-15V Reference Out of Range
E48
Railed Flow Sensor
E49
Blower Pressure Sensor Failure
E50
Flash Memory Failure
E51
Unrecognized Main PCA
E52
Blower Failure
E53
Blower Speed Out of Tolerance
1032443, R
. 00
EV
B
PAP S
I
YNCHRONY
Description of Error
S
& T
ERVICE
ECHNICAL
Device Action
STOP
STOP
STOP
Not Used
STOP
STOP
STOP
STOP
STOP
REBOOT
STOP
STOP
CONTINUE
CONTINUE
REBOOT
Not Used
STOP
REBOOT
STOP
STOP
STOP
STOP
STOP
STOP
STOP
Not Used
STOP
STOP
STOP
T
ROUBLESHOOTING
I
NFORMATION
& D
- P
17
IAGNOSTICS