Invacare® Etude Plus® HC

9 Technical data

9.1 Dimensions

All measurements are stated in cm. All angles are stated in degrees.
All measurements and angles are stated without tolerances.
Invacare®reserves the right to change the stated measurements and
angles.
A
Low
A
219
B
102
C
20–60
D
138–178
The angle between the lower leg section and horizontal is adjustable
from 0º to 15º.
Bed with tilt function

9.2 Mattress dimensions

Side rail:
height
Verso HC
Min. density: 38kg/m
3

9.3 Weights

Max. patient weight
(provided that the weight of the mattress
and the accessories do not exceed 35 kg)
Max. safe working load
(patient + accessories)
Etude Plus standard bed, complete, excluding
accessories
Etude Plus auto-regression, complete, excluding
accessories
Etude Plus bed end with extra plate – 1 piece
20
B
D
C
Etude Plus HC
High
219
102
27–67
145–185
Mattress measures (in cm):
Min.
Max.
Min.
height
width
14
16
90
165 kg
200 kg
77.3 kg
82.7 kg
20.6 kg
Etude Plus bed end without extra plate – 1 piece
Mattress support, standard head end
Mattress support with auto-regression, head end
Mattress support with auto-regression, head end,
without actuator and control box
Mattress support, foot end
Verso II side rail – 1 piece
Lifting pole
Mattress support extension (10 cm)
Mattress support extension (20 cm)
Support handle 40 x 30 cm
Support handle 40 x 40 cm
Swivel handle

9.4 Environmental conditions

Temperature
Relative humidity
Atmospheric
pressure
Be aware that when a bed has been stored under low
temperatures, it must be adjusted to operating conditions
before use.

9.5 Electrical system

Voltage supply: Uin = 120 Volt, AC, 50/60 Hz (AC = Alternating
current)
Maximum current input: Iin max. = 1A
Intermittent (periodic motor operation): Int = max. 10%, 2 min
ON / 18 min OFF
Degree of protection: IPx6* or IPx4** (depending on configuration)
The control unit, external power supply and motors are protected
according to IPx6. When the bed is equipped with a HL80 hand
control, it is protected according to IPx4
Insulation class: Class II
Type B Applied Part
Min.
length
190
Applied Part complying with the specified requirements for
protection against electrical shock according to IEC60601-1.
(An applied parts is a part of the medical equipment which is
designed to come into physical contact with the user or parts that
are likely to be brought into contact with the user.)
Sound level: 45 – 50 dB (A)
* IPX6 classification means that the electrical system is protected
against water projected in powerful jets from any direction.
** IPX4 classification means that the electrical system is protected
against water splashed against the component from any direction.
The bed has no isolator (main switch). If the bed needs to
be electrically disconnected, it has to be unplugged from the
mains socket.
17.9 kg
24.1 kg
27.9 kg
24.1 kg
17.4 kg
7.4 kg
4.2 kg
3.2 kg
5.5 kg
1,6 kg
2,0 kg
1,8 kg
Storage and
Operation
transportation
-10°C to +50°C
+5°C to +40°C
20% to 75%
800 hPa to 1060 hPa
1579562-A