Automatic Defrost (Menu/Tools/Date Time) - Sanyo MIR-554 Instruction Manual

Cooled incubator
Hide thumbs Also See for MIR-554:
Table of Contents

Advertisement

AUTOMATIC DEFROST (MENU/Tools/Date Time)

This product has a frost check opening to check amount of the frost on the evaporator that lowers
temperature in a chamber.
automatically at the specified time everyday. Default setting is 0(Manual).
1. Press the menu button (MENU) to show the menu window and select "Tools" when the top screen is
displayed. Then press the enter key (ENTER).
2. Select "Date Time" on the Select Tools screen, and press the menu button (MENU) to show the menu
window. Select "OK" and press the enter key (ENTER).
S e l e c t
L o g
D a t e
A l a r m , O p t . , L o w
K e y
D e f a u l t
D e l e t e
3. The Date Time screen is displayed. Select the Def Timer (Automatic defrost function) 0 (Manual) or 1
(Auto). When selecting 0 (Manual), automatic defrosting is not done. When selecting 1 (Auto), it is
possible to set the time to defrost automatically in every one hour between 0:00 and 23:00. (The figure
below shows that the defrost starts at 3:00.)
D a t e
D a t e
T i m e
L o g
D e f
CAUTION
Automatic defrosting is activated even in programmed running, and the chamber temperature may arise
due to the amount of the frost existing on the evaporator. In this case, manual defrosting is
recommended.
CAUTION
During defrosting, "Defrosting" is displayed on running status of LCD panel.
During defrosting, the chamber temperature deviates from the set temperature (The chamber
temperature raise by approximately 2
o
30
C and humidity of 80%R.H.)
Care should be taken for deviation of setting and current temperature during defrosting. In case of long
term running of setting temperature below 5
Defrosted water will be evaporated automatically outside of the chamber.
●Defrosting is likely not to operate when the inner temperature is over 5
●Frost will exist on the evaporator when temperature setting is below 5
causes insufficient cooling and inner temperature rising. When a lot of frost existed on the evaporator is
seen through the frost check opening, immediate defrosting is required. Frosting on the evaporator is
accelerated when moisturized material is put inside the chamber.
Automatic defrost function defrosts the frost on the evaporators
T o o l s
M I R - 5 5 4
T i m e
H u m i .
L o c k
P W
S e t t i n g
S e t t i n g
U s e r
D a t a
T i m e
0 8 / 0 6 / 2 0
( Y Y / MM / D D )
1 2 : 1 5 : 0 0
( h h : m m : s s )
I n t e r v a l
6 m i n
T i m e r
1
o
C with the setting temperature of 5
o
C, expected running will be failed due to frost.
M E N U
O K
S v c
S e t t i n g
C a n c e l
( 2 - 3 0 m i n )
( 0 . M a n u a l
1 . A u t o )
3 : 0 0
o
C under ambient temperature of
o
33
C.
o
C. Frost-clogged evaporator

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents