ALARMS AND SAFETY FUNCTIONS

The unit has the alarms and safety function as shown in the table 1 below.
Table 1 Alarms and safety functions
Alarm
1 Auto-set temp.
When the chamber
alarm
temp. deviates from
set temp. by more
than 2.5
5.0
2 Upper limit alarm When the chamber
temp. is higher than
the upper limit temp.
3 Lower limit alarm When the chamber
temp. is lower than
the lower limit temp.
4 Thermal fuse
When the chamber
temp. is over 70
5 Thermal sensor
Input voltage is over
abnormality
70
Input voltage is lower
than -50
corresponding
6 Abnormal
The temp. control
completion of
sensor is higher than
defrosting
20
defrosting
7 Short-circuit of
Microcomputer
SSR or
determines
compressor relay
triac/compressor relay
is shorted
8 Disconnection of
The heater is not
SSR
energized
9 Disconnection of
The compressor relay
compressor relay
is not energized
Condition
<Top screen>
The current chamber
temp. blinks.
o
C.(±1.0∼
o
C changeable)
―――
―――
―――
o
C
<Top screen>
o
E01:Temp. sensor
C corresponding
is opened.
E02:Temp. sensor is
o
shorted.
C
<Top screen>
E03:Abnormal
o
C during
completion of
defrosting.
<Top screen>
E04:SSR or comp.
relay is shorted.
<Top screen>
E05:SSR is
open-circuited.
<Top screen>
E06:Comp. relay is
open-circuited.
Display
Buzzer
Intermittent
tone with
delay
Continuous
tone
Continuous
tone
――
Intermittent
tone
――
Intermittent
tone
Intermittent
tone
Intermittent
tone
45
Remote
Safety operation
contact
Alarm
High stage side;
status
heater OFF
with
Low stage side;
delay
compressor OFF
Alarm
Heater, fluorescent
status
lamp, fan motor
OFF
Alarm
Compressor OFF
status
Main heater and
sub-heater OFF
Alarm
Heater, fluorescent
status
lamp, fan motor
and compressor
OFF
Forced finish of
defrosting
Alarm
――
status
Alarm
――
status
――
Alarm
status