The top screen below is displayed when turning on the power switch. The default temp. is 25.0
date and time are preset at the factory. Refer to page 38 when more accurate setting is needed.
1
S t a n d - b y
T e m p
3
25.0℃
4
5
1. Display of running status
The current running status is displayed. At the power-on, "stand-by" is displayed and the system runs
continuously under the stand-by running condition (refer to page 16).
first time, the system runs continuously under the initial setting condition.
"Running" is displayed at the time of programmed running. "Defrosting" is displayed with reversed and
non-reversed character alternately while removing frost.
"Low Humi." is displayed with reversed and non-reversed character alternately while low humidity mode.
2. Display of program name
A program name under operation is displayed. "Top Screen" is displayed during standby operation.
3. Display of setting
Set value of temperature is displayed.
4. Display of current value
Current value of temperature is displayed.
5. Message display field
A message is displayed when a breakdown occurs.
6. Alarm display(Alarm)
"Alarm " is alternately displayed by reversed/non-reversed character while alarm is operating.
"Warning" is alternately displayed by reversed/non-reversed character at the time of warning.
"Normal" is displayed at a normal condition.
An adding message is displayed in the message display field.
7. Display of door status
"Door" is highlighted when the door is open. "Closed" is displayed when the door is closed.
8. Lighting display (Light)
"ON" is displayed in outline type when the lighting has been turned on. "OFF" is displayed when turned
off. When the lighting is set to be turned on, the system is programmed running, "Program_ON" is
displayed in outline type when the lighting has been turned on and "Program_OFF" is displayed in outline
type when the lighting has been turned off.
9. Display of date and time
The current date and time is displayed.
2
T o p
S c r e e n
o
2 5 . 0
C
2 0 1 0 / 0 1 / 0 1
9
A l a r m
N o r m a l
D o o r
C l o s e d
L i g h t
O F F
1 2 : 0 0 : 0 0
When turning on power for the
15
o
C. The
6
7
8

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents