Sample Absorbance Plate Test Report

Universal Test Plate Analysis
Wavelength: 405
Alignment Results
A1= 0.000 PASS
Accuracy Results
STANDARD
DATA
RESULT
STANDARD
DATA
RESULT
Repeatability Results
READ1
READ2
RESULT
READ1
READ2
RESULT
Turnaround Results
READ1
READ2
RESULT
Spectral Scan Results
580nm= 1.933
581nm= 1.956
582nm= 2.059
583nm= 2.253
584nm= 2.511
585nm= 2.785
Test Plate Standard=587nm
Calculated Peak=586nm
PowerWave Operator's Manual
A12= 0.000 PASS
C01
D04
E02
0.147
2.945
0.618
0.139
2.914
0.613
PASS
PASS
PASS
F12
E09
D11
0.147
2.945
0.618
0.145
2.910
0.613
PASS
PASS
PASS
C01
D04
E02
0.139
2.914
0.613
0.141
2.913
0.616
PASS
PASS
PASS
F12
E09
D11
0.145
2.910
0.613
0.140
2.914
0.614
PASS
PASS
PASS
C01
D04
E02
0.139
2.914
0.613
F12
E09
D11
0.145
2.910
0.613
PASS
PASS
PASS
586nm= 2.892
587nm= 2.753
588nm= 2.516
589nm= 2.334
590nm= 2.186
PASS
Absorbance Plate Test |
H1= 0.000 PASS
H12= 0.000 PASS
F05
G03
H06
2.279
1.133
1.701
2.265
1.128
1.694
PASS
PASS
PASS
C08
B10
A07
2.279
1.133
1.701
2.264
1.125
1.692
PASS
PASS
PASS
F05
G03
H06
2.265
1.128
1.694
2.265
1.128
1.694
PASS
PASS
PASS
C08
B10
A07
2.264
1.125
1.692
2.265
1.126
1.692
PASS
PASS
PASS
F05
G03
H06
2.265
1.128
1.694
C08
B10
A07
2.264
1.125
1.692
PASS
PASS
PASS
45