How To Make And Interpret Srl Measurements; Select Srl As The Type Of Measurement - Agilent Technologies 8712ET User Manual

Option 100 fault location and structural return loss measurements rf network analyzers
Hide thumbs Also See for 8712ET:
Table of Contents

Advertisement

Making SRL Measurements

How to Make and Interpret SRL Measurements

How to Make and Interpret SRL
Measurements
A typical SRL measurement consists of the following steps:

1. Select SRL as the type of measurement.

2. Calibrate the analyzer.
3. Connect the equipment.
4. Determine the connector model.
5. Perform the SRL cable scan.
6. Interpret the SRL measurement.
The next few pages explain how to perform each of these steps.
1. Select SRL as the Type of Measurement
To select SRL mode, press
PRESET
BEGIN
Cable
SRL
.
NOTE
The SRL Cable Scan function operates over a frenquency range of 5 MHz
to 1000 MHz with 1601 measurement points. These values must be used
for all SRL measurements.
Option 100 Fault Location and SRL
4-3