Application-Specific Commissioning; Quick Setup "Commissioning - Endress+Hauser Proline Promass 80 Operating Instructions Manual

Coriolis mass flow measuring system
Hide thumbs Also See for Proline Promass 80:
Table of Contents

Advertisement

Commissioning
!
50
6.3

Application-specific commissioning

In the case of measuring devices without a local display, the individual parameters and functions
must be configured via the configuration program, e.g. ToF Tool - Fieldtool Package. If the
measuring device is equipped with a local display, all the important device parameters for standard
operation can be configured quickly and easily by means of the "Commissioning" Quick Setup
menu.
• Quick Setup "Commissioning", → Page 4 ff.
6.3.1

Quick Setup "Commissioning"

Note!
• The display returns to the cell SETUP COMMISSIONING if you press the ESC key combination
during parameter interrogation. The stored parameters remain valid.
• The "Commissioning" Quick Setup must be carried out before one of the Quick Setups explained
below is run.
• m Only units not yet configured in the current Setup are offered for selection in each
cycle. The unit for mass, volume and corrected volume is derived from the
flow unit.
• n The "YES" option remains visible until all the units have been configured.
"NO" is the only option displayed when no further units are available.
• o Only the outputs not yet configured in the current Setup are offered for selection in
each cycle.
• p The "YES" option remains visible until all the outputs have been parameterized.
"NO" is the only option displayed when no further outputs are available.
• q The "automatic parameterization of the display" option contains the following
basic settings/factory settings:
YES: line 1= mass flow; line 2 = totalizer 1;
information line = operating/system conditions
NO: The existing (selected) settings remain.
Proline Promass 80
corresponding
Endress+Hauser