Application-Specific Commissioning; Commissioning" Quick Setup Menu - Endress+Hauser Proline Promag 51 Operating Instructions Manual

Hide thumbs Also See for Proline Promag 51:
Table of Contents

Advertisement

6 Commissioning
!
76
6.4

Application-specific commissioning

In the case of measuring devices without a local display, the individual parameters and functions
must be configured via the configuration program, e.g. ToF Tool - Fieldtool Package. If the
measuring device is equipped with a local display, all the important device parameters for standard
operation can be configured quickly and easily by means of the "Commissioning" Quick Setup
menu.
6.4.1

"Commissioning" Quick Setup menu

This Quick Setup menu guides you systematically through the setup procedure for all the major
device functions that have to be configured for standard measuring operation.
Note!
In the case of measuring devices without a local display, the individual parameters and functions
must be configured by means of a configuration program, such as ToF Tool - Fieldtool Package from
Endress+Hauser.
ENDRESS+HAUSER
E
-
+
E
ESC
HOME-POSITION
Current Output
Fig. 43:
Quick Setup for commissioning
+ +
Quick Setup
E
+
Freq.-/ Pulse Output
Frequency
Assign
Assign
Current
Frequency
Current
End Value
Span
Freq.
Value
Value
20 mA
f max
Output
Time
Signal
Constant
Failsafe
Time
Mode
Constant
Failsafe
Mode
Automatic parameterization
of the display
Quit Quick Setup
Proline Promag 51
QS
Commission
Language
Defaults
Unit
Volume flow
Measuring
Mode
Quit
Operation
Mode
Pulse
Assign
Pulse
Pulse
Value
Pulse
Width
Output
Signal
Failsafe
Mode
F06-50xxxxxx-19-xx-xx-en-000
Endress+Hauser