Application-Specific Commissioning; Steam Application; Liquid Application - Endress+Hauser Proline Prowirl R 200 Operating Instructions Manual

Hart vortex flowmeter
Hide thumbs Also See for Proline Prowirl R 200:
Table of Contents

Advertisement

Proline Prowirl R 200 HART
1)
Sensor version option "mass (integrated pressure and temperature measurement)", Pressure read in via current input/HART/
Endress+Hauser
4. Setting the write protection switch (WP) on the main electronics module to the ON
position enables hardware write protection. Setting the write protection switch (WP)
on the main electronics module to the OFF position (factory setting) disables
hardware write protection.
 If the hardware write protection is enabled: The Hardware locked option is
displayed in the Locking status parameter . In addition, on the local display the
-symbol appears in front of the parameters in the header of the operational
display and in the navigation view.
If the hardware write protection is disabled: No option is displayed in the Locking
status parameter . On the local display, the -symbol disappears from in front
of the parameters in the header of the operational display and in the navigation
view.
5. Feed the cable into the gap between the housing and main electronics module and
plug the display module into the electronics compartment in the desired direction
until it engages.
6. Reverse the removal procedure to reassemble the transmitter.
10.9

Application-specific commissioning

10.9.1

Steam application

Select medium
Navigation:
Setup → Medium selection
1. Call up the Medium selection wizard.
2. In the Select medium parameter, select the Steam option.
3. When pressure measured value is read in
In the Steam calculation mode parameter, select the Automatic (p-/T-
compensated) option.
4. If pressure measured value is not read in:
In the Steam calculation mode parameter, select the Saturated steam (T-
compensated) option.
5. In the Steam quality value parameter, enter the steam quality present in the pipe.
 Measuring device uses this value to calculate the mass flow of the steam.
Configuring the current output
6. Configure current output→  84.
10.9.2

Liquid application

User-specific liquid, e. g. heat carrier oil
X X X X X X X
X
X
20.50
XX
1)
:
Commissioning
XX
A0029425
127