Figure 3-1:
4-wire cable preparation
Pass the wires through the gland nut and clamping insert.
Gland nut
1. Strip 4-1/2 inch (115 mm) of cable jacket.
2. Remove the clear wrap and filler material.
3. Strip all but 3/4 inch (19 mm) of shielding.
Wrap the drain wires twice around the shield and cut off
Installation Manual
Remove the core processor
Cable glands
Gland supplier
Micro Motion
cable gland
Clamping
insert
Gland type
NPT
1. Strip 4-1/4 inch (108 mm) of cable jacket.
2. Remove the clear wrap and filler material.
3. Strip all but 1/2 inch (12 mm) of shielding.
the excess drain wires.
Drain wires
wrapped around
shield
Go to the shielding
procedure
cover
Cable layout
User-supplied
cable gland
Pass the wires
through the gland.
Terminate the drain
wires inside the
gland.
M20
Wiring
Metal conduit
Run conduit to
sensor
Lay cable in conduit
Done
(do not perform the
shielding procedure)