15-14
SCPI Reference Tables
Table 15-5 (continued)
SENSe command summary
Command
[SENSe[1]]
:FRESistance
:APERture <n> [, <clist>]
:APERture? [<clist>]
:NPLCycles <n> [, <clist>]
:NPLCycles? [<clist>]
:RANGe
[:UPPer] <n> [, <clist>]
[:UPPer]? [<clist>]
:AUTO <b> [, <clist>]
:AUTO? [<clist>]
:DIGits <n> [, <clist>]
:DIGits? [<clist>]
:REFerence <n> [, <clist>]
:STATe <b> [, <clist>]
:STATe? [<clist>]
:ACQuire [, <clist>]
:REFerence? [<clist>]
:AVERage
:TCONtrol <name>
:TCONtrol?
:WINDow <NRf>
:WINDow?
:COUNt <n> [, <clist>]
:COUNt? [<clist>]
[:STATe] <b> [, <clist>]
[:STATe]? [<clist>]
:OCOMpensated <b>
[, <clist>]
:OCOMpensated? [<clist>]
Description
Path to configure four-wire resistance.
Set integration rate in seconds
(3.333333e-5 to 1).
Query aperture integration rate.
Set integration rate in line cycles (60Hz;
0.002 to 60, 50Hz; 0.002 to 50).
Query line cycle integration rate.
Path to set measurement range:
Select range (0 to 120e6).
Query range.
Enable or disable auto range.
Query state of auto range.
Specify measurement resolution (4 to 7).
Query resolution.
Specify reference (0 to 120e6).
Enable or disable reference.
Query state of reference.
Use input signal as reference.
Query reference value.
Path to configure and control filter:
Select filter type: (MOVing or REPeat).
Query filter type.
Set filter window in % of range
(0 to 10).
Query filter window.
Specify filter count (1 to 100).
Query filter count.
Enable or disable filter.
Query state of digital filter.
Enable or disable offset compensation.
Query state of offset compensation.
Model 2701 User's Manual
Default
parameter
Ref
SCPI
Sec 3
(Note 2)
Sec 4
5.0
Sec 4
Sec 4
120e6
ON
7
Sec 4
0
Sec 5
OFF
Sec 4
(Note 3)
0.1
10
(Note 4)
OFF
Sec 3