Storing Readings - Keithley 2701 User Manual

Ethernet-based dmm / data acquisition
Hide thumbs Also See for 2701:
Table of Contents

Advertisement

Model 2701 User's Manual
Perform the following steps to set the date:
1.
2.
3.
Selecting timestamp
Perform the following steps to select either the real-time clock timestamp or the relative
timestamp:
NOTE
1.
2.
3.
4.

Storing readings

Perform the following steps to store readings:
1.
2.
3.
NOTE
4.
5.
NOTE
Press SHIFT and then SETUP.
Use the and
keys to display SET DATE and press ENTER to display the date
in the month/day/year format.
Use the edit keys ( ,
ENTER.
Changing the timestamp will clear the buffer if a storage is in process. The
message "BUF CLEARED" will be displayed to indicate the buffer readings
were lost. If no storage is in process, changing the timestamp will not clear the
buffer.
Press SHIFT and then SETUP.
Use the
and
keys to display "TSTAMP."
Press the
key to place the cursor on the presently selected timestamp (REL or
RTCL).
Use the
or
key to display the relative (REL) or real-time clock (RTCL) and
press ENTER.
Set up the Model 2701 for the desired configuration.
Press the STORE key.
Use the
,
, , and keys to specify the number of readings to store in the
buffer (2 to 450000). Pressing the AUTO key sets the readings count to 000000.
With buffer auto clear disabled, the only valid buffer size values are 450000 and
000000 (which clears the buffer). Any other buffer size value is ignored.
Press ENTER. The asterisk (*) annunciator turns on to indicate the buffer is
enabled. It will turn off when the storage is finished.
The buffer can be stopped at any time by pressing EXIT.
Stored readings are not lost when the instrument is turned off. To clear the
buffer, set the reading value to "000000" and press ENTER.
For remote programming, the continuous storage mode can be selected. After
the buffer fills, operation wraps around to the beginning of the buffer (location
#0) and starts to overwrite old reading data (see TRACe:FEED:CONTrol com-
mand in Table 6-1).
, , and ) to set the date (month/day/year) and press