Echo profile
Notes:
• Double click on each axis to see the Xscale and Data Scale values.
• To zoom in to a section of the profile, left-click and drag a marquee around it. Right-
click inside the window to zoom out.
• Expand or compress the x and/or y axes:
- Left-click on the axis and drag in either direction to reposition the low end of the
scale.
- Right-click on the axis and drag in either direction to reposition the high end of
the scale.
• After saving a profile click on OK, not the x button, to close the Echo Profile Utilities
window, otherwise the profile will not be saved.
•
In the Echo Profile Utilities window click on the tab Echo Profile.
•
Initial profile graph is blank upon entry to dialog. Click on the Measure button to
view and update the profile.
•
It is recommended to use the Detailed resolution view of the echo profile for
troubleshooting. For faster and more coarse views, the Standard resolution may be
used.
•
Click on the Save button and in the new window enter a name and click on OK.
•
Click on OK to exit.
View Saved Echo Profiles
To view a saved profile, click on the tab View Saved Echo Profiles.
Echo Profile Time Based Storage
7ML19985JE06
Resolution
SITRANS LR250 (HART) – INSTRUCTION MANUAL
Page 51