2.3.2. High Calibration Pt.
Distance from sensor reference point
defined in Units (2.2.1.).
Values
Related parameters
When setting the High Calibration Point value, note that echoes are ignored
within Near Range (2.8.1.).
2.3.3. Sensor Offset
A constant offset that can be added to or subtracted from sensor value
compensate if the sensor reference point has shifted. (For example, this could
result from adding a thicker gasket or reducing the standoff/nozzle height.)
Values
Related
parameters

2.4. Rate

Note: Default settings in the parameter tables are indicated with an asterisk (*)
unless explicitly stated.
2.4.1. Response Rate
Note: Changing Response Rate resets Fill Rate per Minute (2.4.2.), Empty
Rate per Minute (2.4.3.), and Damping Filter (2.2.4.).
Sets the reaction speed of the device to measurement changes.
Response Rate
(2.4.1.)
*
Use a setting just faster than the maximum filling or emptying rate (whichever
is faster).
1)
The value produced by the echo processing which represents the distance from sen-
sor reference point to the target. (see
Flanged Horn with extension
14 and
Page 84
SITRANS LR250 (HART) – INSTRUCTION MANUAL
Range: -99.999 to 99.999. Default: 0.00 m
Units (2.2.1.)
Fill Rate per
Minute (2.4.2.)
Slow
0.1 m/min (0.32 ft/min)
Medium
1.0 m/min (3.28 ft/min)
Fast
10.0 m/min (32.8 ft/min)
Threaded Horn Antenna with extension
on page 17).
1)
to High Calibration Point. Units are
Range: 0 to 20 m. Default 0.00 m
Units (2.2.1.)
Near Range (2.8. 1 .)
Empty Rate per
Minute (2.4.3.)
1)
to
Damping Filter
(2.2.4.)
10 s
10 s
0 s
on page
7ML19985JE06