6.1

Reagents

BioTector TOC analyzer uses following reagents:
I. Acid: 1.8 N Sulfuric Acid (H
II. Base: 1.2 N Sodium Hydroxide (NaOH) Reagent
Reagents should not contain high levels of organics, nitrates and phosphates. Ideally, the level of organics,
nitrates and phosphates should be less than 100 g/l (ppb) in the deionized water used to prepare TOC
analyzer reagents.
Acid and Base reagents are stable up to 1 year. Table 7 below summarizes the total days each BioTector
TOC Analyzer reagent lasts at various system analysis range configurations:
Table 7 BioTector TOC Analyzer Reagent Consumption
Container Size
REAGENTS
Acid
Base
Above table is derived from several operation parameters such as 100% online time. Recommended bunds
(reagent spill trays) to contain above quantity reagents are 1x 50 Liters.

Reagents and Calibration Standards

SO
) Reagent containing 80mg/l Manganese Sulfate Monohydrate
2
4
(Liters)
TOTAL DAYS REAGENT LASTS PER ANALYSIS
0-250
mgC/l
25
25
RANGE
0-2000
mgC/l
54
34
53
33
0-20000
mgC/l
32
31