Properties
Electrical information
Operating voltage,
frequency
230 V version
120 V version
Electronic stabilisation of the lamp
voltage (working range: specified
operating voltages ± 10 %)
Power consumption, max.
Fuses, primary
230 V version
120 V version
Protection class
Overvoltage category
Lamp type
Lamp rated voltage
Lamp rated power
Average lamp lifetime
Level 4
Level 5
Level 6
Lighting information
Maximum effective
light guide bundle diameter
Total light flux at light guide exit
(SCHOTT light guide, Ø 15 mm, typ. values)
Level 4
Level 5
Level 6 (max. light flux)
Light entry angle (2α eff )
Without supplementary optics
Heat protection filter
Approvals
230 V version
120 V version
The KL 2500 LCD has been tested and certificated to the applicable standards on electrical
laboratory equipment (EN 61010-1 and UL 3101-1) and to the standard on medical electrical
equipment (EN 60601-1 respective UL 2601-1). This enables manufacturers to obtain easy
approval with integration of the KL 2500 LCD into their medical products.
Die 230 V version features
.
The right is reserved to make changes in the design and supplied items within the scope of on-going technical
development.
Values
220 ... 240 V ~ 50 / 60 Hz
100 V ~ 50 / 60 Hz and
120 V ~ 60 Hz
± 1.5 % of the set lamp
voltage (true RMS value)
W
300
-
T 3,15H,250 V in accordance with
IEC 127-3/5
-
M 6,3 A in accordance with UL198 G
-
II
-
II
-
Halogen reflector lamp
Osram, type HLX 64653
Philips, type 13163
V
24
W
250
h
1500
h
150
h
50
mm
15
lm
500
lm
1000
lm
1300
degrees
approx. 53
-
SCHOTT KG 2, 45 x 45
thickness = 2,0 mm, toughened
-
www.lambdaphoto.co.uk
T: +44 (0)1582 764334
GB
20