Alarms & Safety Functions - Sanyo MCO-5M Instruction Manual

Multi-gas incubator
Hide thumbs Also See for MCO-5M:
Table of Contents

Advertisement

ALARMS & SAFETY FUNCTIONS
This unit has the alarms and safety functions shown in table below, and also self diagnostic functions.
Alarms and safety functions
Alarm & Safety
If the chamber temperature exceeds
Upper limit
the upper limit alarm temperature set
temperature alarm
value.
If the chamber temperature deviates
Automatic set
from the set temperature by ±1
temperature alarm
more.
Automatic set CO
If the chamber CO
2
density alarm
from the set value by ±1% or more.
Automatic set O
If the chamber O
2
from the set value by ±1% or more.
density alarm
When there is no key pressing in
Auto-return
each setting mode for 90 seconds.
Key lock
When the key lock is "ON".
CO
gas cylinder
If the CO
2
empty
when the CO
N
/O
gas cylinder
If the O
2
2
empty
when the N
CO
gas line
When the gas supply line is switched.
2
changeover
(only when MCO-5GC is installed)
O
/N
gas line
2
2
When the gas supply line is switched.
changeover
Wrong CO
gas line
If the CO
2
connection
supply line.
Wrong selection of
If the N
gas between N
and
2
wrongly.
O
2
If the temperature sensor is
Chamber
disconnected.
temperature sensor
abnormality
If the temperature sensor is short
circuited.
29
Situation
o
C or
density deviates
2
density deviates
2
density does not increase
2
gas valve is opened.
2
density does not change
2
/O
gas valve is opened.
2
2
gas is connected to wrong gas
2
gas or O
gas is selected
2
2
Indication
Over heat lamp lights.
E12 or E16 and chamber
Continuous tone
temperature are displayed
alternately.
All digits on the
Intermittent tone with
temperature indicator blink.
15 minutes delay.
All digits on the CO
Intermittent tone with
2
density indicator blink.
15 minutes delay.
All digits on the O
Intermittent tone with
2
density indicator blink.
30 minutes delay.
Normal display mode.
-----
E01 is displayed alternately
with the temperature on the
Intermittent tone
temperature indicator.
E02 is displayed alternately
with the temperature on the
Intermittent tone
temperature indicator.
E01 is displayed alternately
with the temperature on the
temperature indicator.
Intermittent tone
The empty gas supply line
is displayed on the CO
2
density indicator.
E02 is displayed alternately
with the temperature on the
temperature indicator.
Intermittent tone
The empty gas supply line
is displayed on the O
2
density indicator.
E03 is displayed alternately
with the temperature on the
Intermittent tone
temperature indicator.
E04 is displayed alternately
with the temperature on the
Intermittent tone
temperature indicator.
E05 is displayed alternately
with the temperature on the
temperature indicator.
Intermittent tone
E06 is displayed alternately
with the temperature on the
temperature indicator.
85
Buzzer
Safety operation
Heater OFF
Remote alarm
Remote alarm with 15
minutes delay
Remote alarm with 15
minutes delay
Remote alarm with 30
minutes delay
The setting mode is
-----
canceled.
-----
The setting is disabled.
Remote alarm
Remote alarm
Automatic
changeover
of N
/O
gas connection
2
2
port.
Gas supply line is
altered.
Remote alarm
Gas supply line is
altered.
Remote alarm
Remote alarm
N
/O
valve and CO
2
2
Valve are closed.
Remote alarm
N
/O
valve and CO
2
2
Valve are closed.
Heater OFF
Remote alarm