This is the supplement for the original manual (No. IM 01G05B03-01E, 1st Edition) regarding the added and changed
items. Please also refer to this supplement when you read the manual.
Addition to the section "3.5.4 Connection of the Transducers" (page 18) :
(1) Cable Finish
The cable finish should be done as the drawings (a) and (b)
shown in the right.
(2) Cable Gland
When tightening the gland, not only pull back the cable shield
(meshed part) but also roll it out into a ring-shaped form and let
it be pushed by the compression part (grey-colored plastic) so
that the cable shield will have electrical contact with the gland
body. Also, be sure to align the convexity part outside the
compression part and the concavity part inside the gland body.
Refer to the drawings below.
Cable shield rolled out into a ring-shaped form
Compression part
(convexity part outside)
Components of the cable gland
(3) Grounding
The connection of the cable shield will be realized as shown in the drawing below after going through the
procedures (1) and (2) in the above. Connect the main unit to the ground through the terminal "PE" of the power
supply terminals in the main unit. The grounding resistance should be less than 100 ohm.
Transducers
Sensors
(4) Connecting the Sensor ROMs
Connect the sensor ROMs to the corresponding terminals in the main unit when connecting transducers. The
sensor ROMs are packed and affixed to the transducers when shipped from the factory.
Yokogawa Electric Corporation
US300FM
Ultrasonic Flowmeter
Gland body
(concavity part inside)
Junction box
Connection
cable
Connection of the cable shield
(a) Main unit side
Perform
soldering
White
Cable
shield
(meshed)
(*) Finish the brown cables as well as white cables.
(**) You do not connect the thin lines to any terminals.
Gland body
Cap nut
Rubber ring
Compression
part
Cutaway view of the cable gland
Grounding resistance
Main
less than 100 ohm
unit
GND
Manual Change for IM 01G05B03-01E 1st Edition : No.15-010-E July 30 2015-00
Manual Change
No. 15-010-E
(b) Junction box side
Unit : mm
Perform
soldering
Shield line
Shield line
Brown (*)
White
Less
than
10mm
Cable
Thin lines
shield
(**)
(meshed)
Outer skin of the cable
Cable shield
(ringed-shape)
Signal lines
Outer wall of
the main unit
or junction box
Brown (*)
Less
than
5mm
Thin lines
(**)

Advertisement

loading
Need help?

Need help?

Do you have a question about the US300FM and is the answer not in the manual?

Questions and answers

Summary of Contents for YOKOGAWA US300FM

  • Page 1 Connect the sensor ROMs to the corresponding terminals in the main unit when connecting transducers. The sensor ROMs are packed and affixed to the transducers when shipped from the factory. Manual Change for IM 01G05B03-01E 1st Edition : No.15-010-E July 30 2015-00 Yokogawa Electric Corporation...
  • Page 2 Changes to the section “3.1 Scope of Delivery” (page 11) : Change the figure of a nameplate to as below. Addition to the section “3.1 Scope of Delivery” (page 11) : When ordering a couplant weatherproof type (by couplant code “R” in the transducer specification or model code USPA097), a package of “Shin-Etsu Silicone, 1 COMPONENT RTV (KE45T)”...
  • Page 3 The correct explanations for these key operations in the related pages are also as the table below. Related pages: “4.1.1 Interruption of Power Supply” (page 27) “15.2 US300FM doesn’t react anymore” (page 106) Switches the background lighting ON/OFF. LIGHT [Operation during the measurement or menu display]...
  • Page 4 Attention! In case of mass flow, select “Other Medium” as a measured fluid. When “Other Medium” has been selected, US300FM requests to enter the density which is used to calculate mass flow. Manual Change for IM 01G05B03-01E 1st Edition : No.15-010-E July 30 2015-00...
  • Page 5 Changes to the section “7.2 Flow Totalizers” (page 43) : Two of the key operations have been changed as below, requiring pressing the keys “three times” . To reset the two flow totalizers to zero: Press key three times when a totalizer is displayed. To deactivate flow totalizing: Press key three times when a totalizer is displayed.
  • Page 6 Standard Specifications A Standard Specifications US300FM  General Fluid: Measuring Input : Liquid (Turbidity < 10,000 mg/L, Sound velocity 800 to One or two (dual) (Channel A, Channel B). 3,500 m/s, Temperature –30 to +200°C / –22 to +392°F) Current Output: ...
  • Page 7: Operating Conditions

    RS232 serial communication port. is to be connected to US300FM main unit by dedicated (Note) As the RS232 port is on the front panel inside housing, use this function only for test / service / connection cable US300FC.
  • Page 8: Units Of Measurement

    Connection cable is always necessary to connect the shown. Data range for statistical processing can be transducers (junction box) and the US300FM main unit. designated if necessary. The length is specifiable from 1 m (3 ft) to 300 m (984 ft) by the meter.
  • Page 9 Ultrasonic flowmeter Transducers (*3) Model Suffix code Specification Model Suffix code Specification US300FM ····························· Ultrasonic flowmeter US300FT ·························· Transducers for fixed type Usage (*4) -G ····················· General purpose (IP65) Output -A1 ······················ One current output -W ···················· Waterproof (IP67) -A2 ······················ Two current outputs Pipe Size / BG ················...
  • Page 10 (for transducers medium pipe size type, After curing : –40 to +180°C / –40 to 356°F temperature range –30 to +100°C / –22 to Accessories (for ultrasonic flowmeter US300FM) +212°F, set of two magnets) USPA075 Additional magnets for mounting fixture...
  • Page 11: Dimensional Drawing

    Standard Specifications  Dimensional Drawing Ultrasonic flowmeter US300FM Unit: mm (inch) □ □ Transducers US300FT- □ □ □ □ Transducers US300FT- , US300FT- Transducers X m (inch) Y m (inch) X+Y m (inch) Weight kg (lb) US300FT-□B□ 2.0 ( 78.7) 1.0 ( 39.4)
  • Page 12 Standard Specifications Connection Cable US300FC  Wirings Terminal Layout Cable Connection Terminal Designations Terminal Terminal Description Description Name Name Downstream transducer signal for Current output (+,–) channel A Upstream transducer signal for Current output (+,–) when specifying "two channel A current outputs"...