Elcometer 110
®
P.A.T.T.I.
Pneumatic Adhesion Tester
Operating Instructions

Advertisement

Table of Contents
loading

Summary of Contents for Elcometer 110

  • Page 1 Elcometer 110 ® P.A.T.T.I. Pneumatic Adhesion Tester Operating Instructions...
  • Page 2 Class B products are suitable for use in domestic establishments and in establishments directly connected to a low voltage power supply network which supplies buildings used for domestic purposes. is a registered trademark of Elcometer Limited. ® The Elcometer 110 P.A.T.T.I.
  • Page 3: Table Of Contents

    CONTENTS Section Page About your tester ......... . 2 Getting started .
  • Page 4: About Your Tester

    Our products cover all aspects of coating inspection, from development through application to post application inspection. The Elcometer 110 is a world beating product. With the purchase of this product you now have access to the worldwide service and support network of Elcometer.
  • Page 5 See “Accessories and spare parts” on page 18 Figure 1. Elcometer 110 kit of parts The Elcometer 110 Pneumatic Adhesion Tester is packed in a cardboard and foam package. Please ensure that this packaging is disposed of in an environmentally sensitive manner.
  • Page 6: Getting Started

    2 GETTING STARTED 2.1 Basic principle of operation An aluminium pull stub is glued to the test surface. When the glue has fully cured, a pulling piston is attached to the pull stub (Figure 2). A reservoir inside the control module is pressurised by a CO cartridge or by a compressed air line.
  • Page 7: The Control Module

    2.2 The control module Front panel All functions of your tester are controlled from the front panel of the control module (Figure 3). PATTI Piston Pressure (PSIG) Run (Push) 100.0 Piston Rate Reset Supply Pressure Figure 3. Control module - front panel 1.
  • Page 8 Back panel (Figure 4) air-line in zero reset recorder output Elcometer 110 CO cartridge 9V dc in + 44 (0)161 371 6000 sales@elcometer.com www.elcometer.com Serial No Figure 4. Control module - back panel 1. Zero reset: Press the Reset button on the front panel and then, if necessary, rotate Zero reset on the back panel to set the digital piston pressure gauge to zero.
  • Page 9: The Piston

    2.3 The piston Standard pistons Reaction plate Piston hose Piston gasket Connecting hose Piston housing Pull stub Figure 5. Standard piston components, pull stub and hoses F20 Piston The F20 piston (Figure 6) differs from the standard pistons as it has two piston gaskets.
  • Page 10: Power Supply

    2.4 Power supply Your Elcometer 110 Pneumatic Adhesion Tester is powered by a single 9 V dry cell battery or direct from the mains electricity using the optional external power supply. • To fit the battery, slide open the battery compartment at the rear of the tester and insert the battery taking care to observe the correct polarity.
  • Page 11: Preparing For A Test

    4 PREPARING FOR A TEST 4.1 Attaching the pull stub to the coating These instructions describe the use of Araldite epoxy resin to attach the pull stub to the coating. Use of other adhesives may require different surface preparations and/or application techniques. Please refer to the relevant manufacturers recommended procedures.
  • Page 12: Preparing The Control Module

    4.2 Preparing the control module 1. Attach the connecting hose to the piston outlet on the front panel of the control module. 2. Switch the instrument on by rotating the Off/On control. 3. Reset the instrument; press and hold the Reset button, then release. 4.
  • Page 13 supply to the inlet marked ‘air-line in’ on the rear panel of the control module. 2. Open the compressed air supply valve to pressurise the control module. The analogue supply pressure gauge will indicate the pressure. The compressed air supply can be left connected to the control module during testing.
  • Page 14 4.3.3 Pressurising using the trigger inflator (optional item) The trigger inflator is available as an optional accessory - see “Accessories and spare parts” on page 18. 1. Non-threaded cartridges: Unscrew and remove the handle of the trigger inflator and insert the cartridge into the handle with the narrow end up.
  • Page 15: Preparing The Piston

    4.4 Preparing the piston 1. Rotate the Rate control on the front panel of the control module to 0.005. Note: The closed position is 0.000. Do not overtighten the Rate control as this may damage the valve inside the control module. 2.
  • Page 16: Pull Off Test Procedure

    5 PULL OFF TEST PROCEDURE Before you start a test, the adhesive between the pull stub and the coating must be cured. Prepare the instrument and piston - see “Preparing for a test” on page 9. 5.1 Pull off test If the adhesion need only be tested to a specified pull off tensile strength (POTS), the pressure must be determined by the burst pressure (BP) to POTS Conversion Table - see “Piston burst pressure to POTS conversion”...
  • Page 17 4. Unscrew the reaction plate through 90° from the point of contact. This will allow gasket sealing and alignment of the reaction plate perpendicular to the pull stub axis. 5. The digital piston pressure gauge should read 00.0. If not, adjust the zero setting by holding in Reset and at the same time turning the Zero reset control on the back panel of the control module.
  • Page 18: After The Test

    5.2 After the test It is safe to leave the system pressurised between pull off tests. However, if the instrument is to be transported, or left unused for a long period of time, it is advisable to shut down (depressurise) the system. 1.
  • Page 19: Piston Burst Pressure To Pots Conversion

    The tester does not contain any user-serviceable components. In the unlikely event of a fault, the Elcometer 110 should be returned to your local Elcometer supplier or directly to Elcometer. The warranty will be invalidated if the instrument has been opened. Details of Elcometer offices around the world are given on the outside cover of these operating instructions.
  • Page 20: Accessories And Spare Parts

    Connection hose (control module to piston): T11013446 Pure talcum powder: T11013451 Carry case: T11013457 Araldite: T99912906 External power supply, UK, 240 V AC 50 Hz: T99699911 EUR, 220 V AC 50 Hz: T99699912 US, 110 V AC 60 Hz: T99699913...
  • Page 21: Technical Specification

    Internal power supply: 9 V (PP3) External power supply: UK, 240 V AC 50 Hz EUR, 220 V AC 50 Hz US, 110 V AC 60 Hz Display type: 3 digit Liquid Crystal Display Control module accuracy: 0.5 psi Piston accuracy:...
  • Page 22: Related Equipment

    12 RELATED EQUIPMENT In addition to the Elcometer 110, Elcometer produces a wide range of other equipment for determining the physical characteristics of surface coatings. Users of the Elcometer 110 may also benefit from the following Elcometer products: • Elcometer Mechanical Adhesion Testers •...