• Manual mode: push the NIBP Go/Stop hardkey once to
obtain immediate determination.
• Select NIBP from the setup parameters option in Main
Menu to set or change limits or to set up auto mode.
• Auto mode: quick-push the NIBP Auto hardkey to begin
auto BP determinations.
• Stat mode: push-and-hold the NIBP Auto hardkey to
obtain continuous determinations for 5 minutes.
Suspend all monitoring and alarms for NIBP.
Configure a custom auto mode to meet the
needs of your clinical situation.
Set time-cycled (auto mode) NIBP
determinations.
Adjust or auto-set alarm limits. Choose hi to
change the high alarm limit and lo to
change the low alarm limit. You can set
high and low limits individually for systolic,
diastolic, and mean values. The system
maintains appropriate separation between
the high and low limits. Choose auto-set to
use the patient's current readings in setting
the high and low limits.
SpO
2
The SpO
parameter in the Dash 2500 Patient Monitor is
2
available in two different leading technologies: Nellcor
®
MASIMO SET
.
• Select sensor
• Plug sensor connector into SpO
• Apply sensor
• Select SpO
from the setup parameters option in the
2
Main Menu
• Set up monitoring parameters
NOTE: Attach the SpO
sensor to the limb opposite the one
2
with the blood pressure cuff.
NOTE: Remove nail polish and artificial nails.
3
Quick guide-new.indd 4
®
and
port
2
Suspend all monitoring and alarms for SpO
Choose whether or not the plethysmographic
waveform appear on the screen.
Adjust or auto-set alarm limits. Select hi
to change the high alarm limit and lo to
change the low alarm limit. The system
maintains appropriate separation between
the high and low limits.
NOTE: The auto-set option is not available for SpO
choose the auto-set option to set the alarm limits for oxygen
saturation. However, when SpO
2
auto-set in the Setup HR/Pulse menu will set the alarm limits
around the SpO
PR value.
2

Heart Rate

Heart rate/pulse (HR/Pulse) in the Dash 2500 Patient
Monitor can be derived from one of several sources. In
descending order of priority, they are ECG, pulse oximetry
(SpO
), and non-invasive blood pressure (NIBP). Instead of
2
choosing a single source of HR/Pulse, you can activate the
auto feature to choose the best available source.
• Prepare patient
• Connect monitoring lines
• Select HR/Pulse from the setup parameters option in
the Main Menu
• Set up monitoring parameters
Choose the source of the HR/Pulse data to
be displayed on the screen.
Adjust the volume of the tone that sounds
each time a QRS complex is detected.
Adjust or auto-set alarm limits. Select hi
to change the high alarm limit and lo to
change the low alarm limit. The system
maintains appropriate separation between
the high and low limits. Choose auto-set to
use the patient's current readings in setting
the high and low limits.
2
. You cannot
2
is the source for HR/Pulse, using
.
12-7-17
9:33