Electrical Installation; Installation In Conduit - Hach sc100 1720E Instrument Manual

Analysis system
Table of Contents

Advertisement

Installation
3.2

Electrical Installation

DANGER
The instrument must be installed
by qualified technical personnel
for adherence to all applicable
electrical codes. The 1720E/sc100
product configuration is not
intended for installation in
hazardous locations.

3.2.1 Installation in Conduit

3.2.2 Installation Using a Power Cord
DANGER
Use of power cords is not
permitted in hazardous locations.
Figure 12
Removing Voltage Barrier
J1
J1
S1
S1
J4
J4
NETWORK
NETWORK
INTERFACE
INTERFACE
CARD
CARD
1.
High voltage barrier
14
High-voltage wiring for the controller is conducted behind the high voltage
barrier in the controller enclosure. The barrier must remain in place unless a
qualified installation technician is installing wiring for power, alarms, or relays.
See
Figure 12
In hard-wired electrical applications, the power and safety ground service drops
for the instrument must be 18 to 12 AWG. See
relief and conduit opening sealing plug information. See
for wiring information.
Where permitted by local electrical codes, a sealing-type strain relief to
maintain the NEMA 4X/IP66 environmental rating and a power cord less than
3 meters (10 feet) in length with three 18-gauge conductors (including a safety
ground wire) can be used, see
See
Figure 13 on page 15
assembly. See
J3
J3
J2
J2
U5
U5
J5
J5
U9
U9
J6
J6
for barrier removal information.
Replacement Parts and Accessories on page
for strain relief and conduit opening sealing plug
section 3.2.3 on page 15
PCB
PCB
PCB
PCB
CONNECTOR
CONNECTOR
CONNECTOR
CONNECTOR
2.
Unsnap the barrier latch then pull out to remove the barrier.
Figure 13 on page 15
section 3.2.3 on page 15
for wiring information.
1
1
+
+
+
+
DATA
DATA
DATA
DATA
2
2
– DATA
– DATA
3
3
SERVICE REQUEST
SERVICE REQUEST
4
4
+
+
+
+
V
V
V
V
5
5
GND
GND
6
6
1
1
+
+
+
+
OUT 2
OUT 2
OUT 2
OUT 2
2
2
– OUT 2
– OUT 2
3
3
SHIELD/CHASSIS GND
SHIELD/CHASSIS GND
4
4
+
+
+
+
OUT 1
OUT 1
OUT 1
OUT 1
5
5
– OUT 1
– OUT 1
FIELD WIRING
FIELD WIRING
FIELD WIRING
FIELD WIRING
INSULATION MUST
INSULATION MUST
INSULATION MUST
INSULATION MUST
BE RATED TO
BE RATED TO
BE RATED TO
BE RATED TO
80° C MINIMUM
80° C MINIMUM
80° C MINIMUM
80° C MINIMUM
for strain
55.
1
2