TEST WARNING: This message will be displayed if ESC was pressed
during testing. Lines 1 and 2 will flash prompting the user to select
ENTER to continue test, or ESC to end test. This message will hold for 5
seconds and then revert back to its respective screen. If the test is
cancelled, a new cassette will be required to repeat the test.
Patient Data Record – This record is displayed over three screens. Each
screen can be accessed by using the ← and → keys. The complete
record will be printed automatically if AUTOPRINT is set to ON, or it may
be printed/reprinted by pressing the PRINT/ENTER key.
Patient results are POSITIVE, NEGATIVE, or INVALID.
An INVALID result should be repeated. (See Section 7, Item 13.)
Invalid results will not be stored in memory.
NOTE: Internal Controls are performed automatically during each Rapid
fFN test. These internal controls check for (1) a threshold level of
signal at the procedural control line, (2) proper sample flow
across the Rapid fFN Cassette, (3) absence of conjugate
aggregation, and (4) proper functioning of the TLi
hardware.
TLi
System User Manual
IQ
Section 4 − Software Functions - Detailed Descriptions
WANT TO CANCEL TEST
FETAL FIBRONECTIN
03:00 PM
PT:JANE DOE 123
RESULT:POSITIVE
INTERNAL CONTROLS
USER:JOHN SMITH
ANALYZER: PASS
CASSETTE: PASS
CAL CODE:FG56
CASSETTE LN:L1002
ANALYZER ID:00426
analyzer
ESC-MAIN MENU
IQ
ARE YOU SURE YOU
ENTER-CONTINUE
ESC-MAIN MENU
10/26/07
← →
4 − 9