The CASSETTE LOT# must be entered to proceed to the next step. The
CASSETTE LOT# is located on the cassette pouch. The software
requires that the lot number is entered in the correct format: one alpha
character followed by four numeric characters (e.g., L1002).
The CALIBRATION CODE# must be entered to proceed to the next step.
The CALIBRATION CODE# is located on the cassette box. The software
requires that the code number is entered in the correct format: two alpha
characters followed by two numeric characters (e.g., FG56).
NOTE: The calibration code is established by Hologic for each lot of
Rapid fFN Cassettes.
Calibration Data Record - This record is displayed over two screens.
Each screen can be accessed by using the → and ← keys. The
complete record will be printed automatically if AUTOPRINT is set to ON,
or it may be printed/reprinted by pressing the PRINT/ENTER key.
TLi
System User Manual
IQ
Section 4 − Software Functions - Detailed Descriptions
SET CALIBRATION
CASSETTE LOT#
>L1002
ENTER-ACCEPT
SET CALIBRATION
CALIBRATION CODE#
>FG56
ENTER-ACCEPT
FETAL FIBRONECTIN
03:00 PM
10/26/07
SYSTEM CALIBRATED
USER:JOHN SMITH
CAL CODE:FG56
CASSETTE LN:L1002
ANALYZER ID:00426
ESC-MAIN MENU
4 − 4