TEST PATIENT (Internal Incubation Mode)
1. From the Main Menu, press 1 to select TEST PATIENT.
2.
Enter USER ID and press ENTER.
3. Enter the last two digits of the CASSETTE LOT# (on cassette
pouch) and press ENTER. The lot number must be entered to
proceed to the next step.
NOTE: The analyzer will automatically compare the CASSETTE LOT#
used to set calibration with the cassette lot number used for
patient testing. If the lot numbers do not match, the analyzer will
request the user to recalibrate the system. The cassette lot
number used for calibration will be displayed on the third line.
TLi
System User Manual
IQ
Section 3 – General Operating/Testing Instructions
fFN MAIN MENU
1-TEST PATIENT
2-SET CALIBRATION
3-DAILY QC
TEST PATIENT
USER ID
>JOHN SMITH
ENTER-CONFIRM
TEST PATIENT
CASSETTE LOT#
>L1002
ENTER-ACCEPT
3 − 4
↓