After instrument installation, the TLi
Read Section 4 for detailed descriptions of displays, prompts and operating sequences.

STARTING THE SYSTEM

1. Turn on the analyzer using the on/off switch located on the left side
of the analyzer.
The screen will display SYSTEM DIAGNOSTICS. If the analyzer fails
the self-test, two beeps will sound; otherwise the analyzer will go to the
next screen.
Once SYSTEM DIAGNOSTICS is complete, the display will change to
the software VERSION and the DATE and TIME for five seconds and
then to the Main Menu. Verify the date and time are correct.
See Section 4 for setting Date/Time.
2. Turn on the printer using switch at rear and ensure labels
are in the printer. See Section 6 for loading printer labels.
TLi
System User Manual
IQ
Section 3 – General Operating/Testing Instructions
SECTION 3 – GENERAL OPERATING / TESTING INSTRUCTIONS
analyzer can be operated on a day-to-day basis by using the following procedures.
IQ
HOLOGIC TLi SYSTEM
VERSION 2.0
SYSTEM DIAGNOSTICS
IN PROCESS
HOLOGIC TLi SYSTEM
VERSION 2.0
03:00 PM
10/26/07
3 − 1